Global Global Algérie Français Algeria English Angola Português Angola English Argentina Español Argentina English Australia English Austria Deutsch Austria English Azerbaijan English Azerbaijan Русский Bahrain English Bangladesh English Belgium English Belgium Français Belgium Nederlands Brazil Português Brazil English Bulgaria български Bulgaria English Cameroon English Cameroon Français Canada English Canada Français Chile Español Chile English China 中文 China 日本語 China English Colombia Español Colombia English Croatia English Czech Republic Čeština Czech Republic English Denmark English Djibouti English Ecuador Español Ecuador English Egypt English Estonia English Ethiopia English Finland Suomi Finland English France Français France English Georgia English Georgia Русский Germany Deutsch Germany English Ghana English Greece Ελληνικά Greece English Guatemala Español Guatemala English Hong Kong, China English Hong Kong, China 中文 Hungary Magyar Hungary English India English Indonesia English Iraq English Ireland English Italy Italiano Italy English Ivory Coast Français Ivory Coast English Japan 日本語 Japan English Jordan English Kazakhstan Русский Kazakhstan English Kenya English Kuwait English Kyrgyzstan Русский Kyrgyzstan English Latvia English Lithuania English Malaysia English Mauritius English Mauritius français (Maurice) Mexico Español Mexico English Moldova Română Moldova English Mongolia English Morocco English Morocco Français Mozambique Português Mozambique English Netherlands Nederlands Netherlands English New Zealand English Nigeria English Norway English Oman English Pakistan English Paraguay Español Paraguay English Peru Español Peru English Philippines English Poland Polski Poland English Portugal Português Portugal English Qatar English Republic of Korea 한국어 Republic of Korea English Romania Română Romania English Saudi Arabia English Serbia Српски Serbia English Singapore English Slovakia English Slovenia English South Africa English Spain English Spain Español Sri Lanka English Sweden English Switzerland Deutsch Switzerland Français Switzerland Italiano Switzerland English Tanzania English Thailand ไทย Thailand English Togo English Togo Français Tunisia English Tunisia Français Türkiye Türkçe Türkiye English Turkmenistan Русский Turkmenistan English Ukraine Українська Ukraine English United Arab Emirates English United Kingdom English Uruguay Español Uruguay English USA English Uzbekistan English Uzbekistan Русский Vietnam Tiếng Việt Vietnam English

Low-temperature demagnetization curve testing evaluates the second-quadrant magnetic performance of permanent magnets such as NdFeB, SmCo, and ferrite under cryogenic and sub-zero service conditions, providing critical data on remanence, coercivity, and irreversible loss behavior. This article explains the testing principles, sample requirements, measurement methods, applicable standards, and typical applications of low-temperature demagnetization curve experiments performed by accredited third-party laboratories.

Technical Overview

The demagnetization curve is the second-quadrant portion of the hysteresis loop of a permanent magnet, describing how magnetic flux density (B) and magnetic polarization (J) decay as the applied reverse field (H) increases. It is the single most important dataset for magnet selection and magnetic circuit design, because it defines whether a magnet can maintain its operating point on a given load line without irreversible demagnetization.

A low-temperature demagnetization curve experiment measures this curve at controlled temperatures below ambient, typically from 0 °C down to -196 °C (liquid nitrogen temperature). The key parameters extracted at each temperature point include:

  • Remanence (Br): the residual flux density at zero applied field.
  • Coercive force (Hcb): the reverse field at which flux density B reaches zero.
  • Intrinsic coercivity (Hcj): the reverse field at which magnetic polarization J reaches zero, representing the true resistance to demagnetization.
  • Maximum energy product ((BH)max): the peak value of the B×H product on the curve.
  • Knee point (Hk) and squareness ratio: the field at which the curve begins to bend, often defined at J = 0.9 Br, and the ratio Hk/Hcj describing curve rectangularity.

Low-temperature behavior is non-trivial and material-dependent. For sintered NdFeB, remanence increases as temperature falls (temperature coefficient of Br is approximately -0.09 to -0.13 %/K around room temperature), but intrinsic coercivity exhibits a well-documented low-temperature anomaly: below roughly 200 K, Hcj decreases rather than continues to rise, increasing the risk of irreversible demagnetization in cryogenic or arctic service. Sintered SmCo magnets retain more predictable coercivity at low temperature, while hard ferrite magnets suffer significant coercivity loss in the cold, and AlNiCo magnets, with their very low coercivity, must be designed with high permeance coefficients regardless of temperature. Reliable low-temperature curve data is therefore essential for design margins.

Measurements are performed in accordance with recognized standards, including IEC 60404-5 (measurement of magnetic properties of permanent magnets by the hysteresisgraph method), IEC 60404-8-1 (specifications for hard magnetic materials), ASTM A977/A977M (magnetic properties of permanent magnet materials at elevated and low temperatures using a vibrating sample magnetometer), and the corresponding national standards such as GB/T 3217 for hysteresisgraph measurement of permanent magnets.

Testing Samples

Sintered NdFeB Magnets

Sintered neodymium-iron-boron magnets are the most frequently submitted specimens because of their strong temperature sensitivity. All commercial grade families (N, M, H, SH, UH, EH, and AH series) may be characterized at low temperature, with high-coercivity Dy- or Tb-containing grades commonly tested for automotive and aerospace qualification. Specimens are typically machined blocks, discs, or cylinder segments with two parallel, ground faces for pole contact in closed-circuit measurement.

Sintered SmCo Magnets

SmCo5 and Sm2Co17 type magnets are tested for cryogenic and space applications where stable coercivity and low irreversible loss are required over wide temperature ranges, including excursions to -196 °C. These specimens are often smaller in cross-section but require very high saturation fields due to their large anisotropy fields.

Bonded and Injection-Molded Magnets

Bonded NdFeB and bonded ferrite magnets, produced by compression bonding or injection molding with polymer binders, are tested to verify low-temperature performance for sensors, stepper motors, and micro-motor applications. The binder must be verified to remain mechanically intact at the test temperature.

Ferrite and AlNiCo Magnets

Hard ferrite magnets used in automotive sensors and motors require low-temperature verification because their coercivity decreases significantly in the cold. AlNiCo specimens, with low Hcb and Hcj, must be measured with careful attention to shape permeance and are usually tested in magnetized assemblies or with high length-to-diameter ratios.

Magnet Assemblies and Motor Components

Rotor and stator assemblies, magnetic couplings, and complete device subassemblies may be evaluated for flux output and irreversible loss after low-temperature exposure using open-circuit flux or moment measurements rather than full B-H curves.

Sample Preparation Requirements

  • Specimens must be fully saturated before measurement, typically by a pulsed field or superconducting magnet with sufficient field strength (up to 10 T or more for high-Hcj grades at low temperature).
  • Geometry should provide a well-defined demagnetization factor; for closed-circuit hysteresisgraph testing, two flat, parallel, and reasonably smooth faces are required for pole contact.
  • Specimens should be clean, dry, and free of coatings that could become brittle or crack during thermal cycling unless the coating itself is part of the evaluation.
  • Sufficient thermal soak time (typically 30 to 60 minutes depending on specimen mass) must be allowed at each set point so the magnet core reaches equilibrium before data acquisition.

Testing Items

A typical low-temperature demagnetization curve experiment reports the following items at each specified temperature:

  • Full second-quadrant B(H) demagnetization curve and intrinsic J(H) curve.
  • Br, Hcb, Hcj, and (BH)max values at each temperature point.
  • Knee field Hk (commonly at J = 0.9 Br) and squareness ratio Hk/Hcj.
  • Temperature coefficients of Br, Hcb, and Hcj derived between reference points (for example, between +20 °C and -40 °C).
  • Reversible and irreversible flux loss after cooling and rewarming cycles.
  • Recoil line behavior and residual flux density after partial demagnetization at temperature.
  • Optional aging or thermal cycling sequences with post-cycle verification at room temperature.

Common test temperature points include +20 °C (reference), 0 °C, -20 °C, -40 °C (automotive qualification), -55 °C (military/aerospace), -80 °C, -110 °C, and -196 °C (cryogenic, liquid nitrogen). Custom set points are defined by the customer's service specification.

Testing Methods

Closed-Circuit Hysteresisgraph with Climatic or Cryogenic Chamber (IEC 60404-5)

The reference method places the specimen between the pole pieces of a high-permeability electromagnet fitted with an environmental chamber or cryostat. B and J detecting coils (or a combination of a search coil and a Hall probe) surround or contact the specimen, and the applied field H is swept from positive saturation through the second quadrant. The chamber controls temperature between approximately -196 °C and +200 °C with typical stability of ±1 °C or better. Calibration is performed using reference magnets or nickel reference specimens traceable to national metrology institutes, and pole-piece saturation is verified to remain negligible. This method gives the most accurate absolute values for appropriately sized specimens.

Vibrating Sample Magnetometer at Low Temperature (ASTM A977/A977M)

Small specimens (typically a few milligrams to a few grams) are measured in a VSM equipped with a cryostat, often with a superconducting solenoid providing fields up to 7 to 9 T or higher to fully saturate high-anisotropy grades at low temperature. Measured magnetic moment curves are corrected for the specimen demagnetization factor to reconstruct the intrinsic J(H) curve. This method is ideal for material development, grain-boundary engineering studies, and cryogenic characterization of grades whose required saturation field exceeds conventional electromagnets.

Pulsed-Field Magnetometry

Pulsed-field magnetometers apply capacitor-discharge fields exceeding 10 T, enabling saturation and full curve extraction for the highest-coercivity EH and AH grades at low temperature. Because the pulse is fast, eddy-current effects in electrically conductive specimens must be modeled or minimized through specimen segmentation and slower pulse options.

Open-Circuit Flux and Moment Measurements

For assemblies and magnets measured on a Helmholtz coil with a fluxmeter, or on a moment-measuring coil set, the magnetic flux or moment is recorded after saturation and after exposure to each low-temperature set point, including any applied reverse bias field. This approach quantifies irreversible loss against the magnet's operating permeance coefficient and is widely used for qualification of motor rotors and complete devices.

General Test Procedure

  • Verify specimen identity, dimensions, and surface condition; record mass and geometry.
  • Demagnetize and then fully saturate the specimen; confirm saturation by a second identical pulse producing no change in moment.
  • Stabilize the chamber at the first temperature set point and soak until the specimen core reaches equilibrium.
  • Acquire the B(H)/J(H) curve by sweeping the applied field through the second quadrant to the maximum specified field.
  • Repeat at each set point, either on a single specimen in sequence (lowest risk of thermal shock first) or on dedicated specimens per point.
  • For flux-loss studies, re-measure at room temperature after each cold exposure to separate reversible from irreversible changes.
  • Issue a report with all curves, parameters, temperature points, standards applied, and measurement uncertainty.

Applications

  • Electric vehicle traction motors and actuators: verifying that magnet grades retain sufficient Hcj and knee margin at -40 °C cold-start and cold-cruise conditions, including transient reverse fields from fault currents.
  • Aerospace and satellite systems: qualifying magnets for thermal cycling between launch, orbital, and deep-space environments, often down to -196 °C or lower.
  • Cryogenic machinery: magnetic bearings, couplings, seals, and cryopumps operating near liquid nitrogen (77 K) or liquid hydrogen (20 K) temperatures.
  • Superconducting and MRI systems: room-temperature and cryogenic shimming magnets, field-cancellation assemblies, and accessory magnets exposed to cold surfaces.
  • Wind energy and outdoor power equipment: generators and sensors installed in arctic or high-altitude cold climates.
  • Supplier qualification and incoming inspection: automotive and industrial specifications increasingly require certified demagnetization curves at temperature as part of PPAP-style documentation, executed by accredited third-party laboratories.
  • Materials research: quantifying the low-temperature coercivity anomaly of NdFeB and evaluating mitigations such as heavy-rare-earth additions, grain-boundary diffusion, and grain refinement.

FAQ

Why does the intrinsic coercivity of NdFeB decrease at low temperature?

Sintered NdFeB coercivity is controlled by nucleation and domain-wall motion at grain boundaries. Although the magnetocrystalline anisotropy field increases as temperature falls, the actual coercivity reaches a maximum near room temperature and then declines at cryogenic temperatures, a behavior linked to changes in domain-wall energy and pinning at the grain-boundary phase. This anomaly cannot be predicted by linear extrapolation of high-temperature coefficients, which is why direct low-temperature measurement is necessary.

Which temperatures should be tested?

Temperature points should bracket the worst-case service condition of the device. Typical selections are -40 °C for automotive and outdoor equipment, -55 °C for many aerospace and defense programs, and -110 °C to -196 °C for cryogenic and space applications, always with a +20 °C reference measurement for comparison.

What is the difference between the B(H) and J(H) curves?

The B(H) curve describes the actual flux density in the magnet including the field contribution, and its intersection with zero defines Hcb. The J(H) intrinsic curve removes the vacuum field term and describes the magnetization itself; its intersection with zero defines Hcj. Permanent magnet designers primarily use the knee of the J(H) curve to judge irreversible demagnetization risk.

How are pass/fail criteria defined?

The usual criterion is that, at the lowest service temperature and maximum reverse field (including fault transients), the magnet's operating point defined by its load line must remain above the knee of the intrinsic curve with a specified safety margin. Alternatively, specifications impose a minimum Hcj or a maximum allowable irreversible flux loss at temperature.

How long does the experiment take?

A single temperature point with soaking typically requires 1 to 2 hours, and a multi-point profile (for example, five temperatures from +20 °C to -196 °C) with reporting generally takes two to five working days, depending on specimen count, saturation requirements, and thermal cycling sequences.

Is the test destructive?

Closed-circuit hysteresisgraph and open-circuit flux measurements are non-destructive, and specimens can usually be returned. VSM testing requires cutting a small specimen from the part and is therefore consumed in preparation. Coated or adhesive-bonded assemblies may exhibit coating or bond effects during extreme thermal cycling.

What should a compliant test report contain?

A compliant report includes specimen identification and geometry, magnet grade and lot, test standards applied (such as IEC 60404-5 or ASTM A977/A977M), temperature set points and stability, saturation field used, plotted B(H) and J(H) curves, tabulated Br, Hcb, Hcj, (BH)max, Hk and squareness at each temperature, derived temperature coefficients, flux-loss results if applicable, calibration traceability, and stated measurement uncertainty, issued under the scope of an accredited third-party laboratory.